BA 194C/07/en/07.02
No. 51500268
Software version 2.50 or later
Liquisys M
CPM 223 / 253
Transmitter for pH and Redox
Endress Hauser
The Power of Know How
Quality made by
Endress+Hauser
ISO 9001
ENDRESS+HAUSER
LIQUISYS S
ENDRESS+HAUSER
LIQUISYS S
Operating Instructions
slp 8.0 deen 220802
Zobrazit stránku 0
1 2 3 4 5 6 ... 91 92

Shrnutí obsahu

Strany 1 - Endress Hauser

BA 194C/07/en/07.02No. 51500268Software version 2.50 or laterLiquisys MCPM 223 / 253Transmitter for pH and RedoxEndress HauserThe Power of Know HowQua

Strany 2

3.3 Mounting3.3.1 Field instrumentSeveral mounting options are available for theLiquisys M in the field instrument version:• Post mounting on cylindri

Strany 3 - Table of contents

Post mounting kitMounting kit for installation of field housing onhorizontal and vertical pipes (max. ∅ 60 mm);material: stainless steel SS 304 (AISI

Strany 4 - 1 General information

3.3.2 Mounting examplesMONT10.CDRFig. 3.7Pipe mounting ofLiquisys M field instrumentMONT2.CDRFig. 3.8Liquisys M field instrumentleft:Wall mountingrigh

Strany 5 - 1.5 Product structure

3.3.3 Panel-mounted instrumentThe device is attached using the supplied tensioning screws (see Fig. 3.9).The required overall installation depth is ap

Strany 6 - 2 Safety

3.4 Electrical connectionConnection diagramThe connection diagram depicted in Fig. 3.10 shows the connections of an instrumentequipped with all the op

Strany 7 - 2.6 Declaration of conformity

Note:• Please label the sensor terminalblock with the enclosed sticker.• Connect the ground terminal withPE.Caution:• Terminal designated as NCmay not

Strany 8 - 3 Installation

3.5 Electrode installation and measuring cable connectionMeasuring cable connectionThe terminal block is located in a separateconnection compartment w

Strany 9 - 3.2 Dimensions

Cable termination for panel-mounted instrument CPM 223 with solder-free BNC plug(for glass electrodes only)1. Cut off wire end sleeves of the coaxialc

Strany 10 - 3.3 Mounting

Note:• The instrument is pre-configuredfor symmetrical measurement.Change the configuration in fieldA2 for asymmetrical measurement(see chapter 5.2.1)

Strany 11 - PM253E03.CHP

4 Operation4.1 Operator interface4.2 DisplayLED indicatorsLiquid crystal display705pHR262SetpointENDRESS+HAUSERLIQUISYS MDISPLAY1.CDRLiquid crystal di

Strany 12 - 3.3.2 Mounting examples

Need information on the instrument?Please read the following chapters:SafetyGeneral informationYou wish to install the instrument.The required steps a

Strany 13 - Endress+Hauser 11

4.3 KeysCAL keyWhen the CAL key is pressed, the instrument prompts for the calibrationaccess code:Code 22 for calibrationCode 0 or any code for calibr

Strany 14 - 3.4 Electrical connection

4.4 Auto / manual mode of operationNote:• Enable the manual mode byentering access code “22”.• The operating modes remains ineffect even after a power

Strany 15 - 131094-4G

4.5 Operating concept4.5.1 Operating modesNote:• Remains in Setup mode forapprox. 15 mins., the systemautomatically jumps back tomeasuring mode. An ac

Strany 16 - Fig. 3.13

4.5.2 Access codesAll instrument access codes are fixed, i.e. they cannot be modified. When the instrumentrequests the access codes, it recognises the

Strany 17

4.5.4 Hold function: “freezing” of outputsfreezes” the outputsThe current output is “frozen” in the setup mode and during calibration, i.e. the last c

Strany 18 - • If the software setting

5 Instrument configurationAfter power-up the instrument performs aself-test and then enters to measuring mode.Now it can be configured and calibrated

Strany 19 - Setpoint

Factory settingsWhen the instrument is switched on for the first time, the factory settings are in effect. Thefollowing table provides an overview of

Strany 20 - 4.3 Keys

5.1 Start-upAfter switching the instruement on, make the following settings to the specified function groups:• Function group SERVICE (S)S1: Select la

Strany 21 - PM253ED4.CHP

5.2.2 Setup 2 (Temperature)Coding FieldSelection or rangeFactory setting (bold)Display InfoBFunction groupSETUP 2Initial display in function groupSETU

Strany 22 - 4.5 Operating concept

5.3 Current inputThis function group offers two independent application solutions, provided that the currentoutput of an external measured quantity, e

Strany 23 - • See the appendix of these

PM253E00.CHPTable of contents1.1 Symbols used ... 21.2 Storage and transport ...

Strany 24

5.3.2 Feedforward control to PID controllerIn processes with very short response times it may be practical to apply the flow rate to thecontroller, if

Strany 25 - 5 Instrument configuration

Coding FieldSelection or rangeFactory setting (bold)Display InfoZFunction groupCURRENT INPUTInitial display infunction groupCURRENT INPUT.Z1Select flo

Strany 26 - RELAY2.CDR

5.4 Current outputsThe function group CURRENT OUTPUT is used to configure the individual outputs. Either alinear (O3 (1)) or, in conjunction with the

Strany 27 - Oper.Mode

Coding FieldSelection or rangeFactory setting (bold)Display InfoOFunction groupCURRENTOUTPUTInitial display in function groupCURRENT OUTPUT.O1Selectio

Strany 28 - Glass A5

Coding FieldSelection or rangeFactory setting (bold)Display InfoO3 (2)Current outputsimulationlin = linear (1)sim = simulation (2)Tab=Tabelle (3)The s

Strany 29 - Alarm signalling and

5.5 Monitoring functionsThe monitoring functions are used to define various alarms and to set output contacts. Eachindividual error can be defined to

Strany 30 - 28 Endress+Hauser

Coding FieldSelection or rangeFactory setting (bold)Display InfoF6Set alarm contactto be effective forselected erroryesnoIf set “no” all the other ala

Strany 31 - PM253E5A.CHP

5.5.2 CheckThe function group CHECK is only accessible for instruments equipped with the Plus packet.Two different monitoring functions can be selecte

Strany 32 - User-defined current output

Limit monitoringThis function monitors the measuring value for exceeding monitoring limits and triggers analarm.Coding FieldSelection or rangeFactory

Strany 33 - Sel. Out2

Coding FieldSelection or rangeFactory setting (bold)Display InfoP8Enter upper alarmthreshold16.00 pH-2.00 ... 16.00 pHOmitted when P5 = Off.P9Select p

Strany 34 - Table (3)

1 General information1.1 Symbols usedWarning:This symbol alerts to hazards whichcould cause serious injuries as wellas damage to the equipment ifignor

Strany 35 - 5.5 Monitoring functions

5.6 Relay contact configurationThe function group RELAYS is only accessible for instruments equipped with the Plus packet.The relay contacts described

Strany 36 - CleanTrig

5.6.2 P(ID) controllerThe Liquisys M supports the definition of various controller functions. On the basis of the PIDcontroller, P, PI, PD and PID con

Strany 37 - • Due to its semiconductor

Control and fine-optimisation of set parameters using a recorderActuating signal outputs (R237 ... R2310)The control contact in question outputs a swi

Strany 38 - LowAlarm

Control characteristic for direct and inverted control actionField R236 offers two control characteristics for selection which have the effects shown

Strany 39 - Tmax High

5.6.4 ChemoClean functionJust like the timer function, ChemoClean canalso be used to start a cleaning cycle.However, ChemoClean supports differentclea

Strany 40 - GRENZ-E.CDR

Coding FieldSelection or rangeFactory setting (bold)Display InfoRFunction groupRELAISRelay contacts can be selectedand adjusted.R1Selection ofcontact

Strany 41 - not allowed

Coding FieldSelection or rangeFactory setting (bold)Display InfoR216Enter alarmthreshold(as an absolutevalue)pH 16.00pH –2,00 ... 16.001500 mV–1500 ..

Strany 42 - Pulse frequency modulation

Coding FieldSelection or rangeFactory setting (bold)Display InfoR226Enter alarmthreshold(as an absolutevalue)150.0 °C–50.0 ... 150.0 °CWhen the alarm

Strany 43 - PM253E5B.CHP

Coding FieldSelection or rangeFactory setting (bold)Display InfoR236Selection ofcharacteristiccontrollerdir = directinv = invertedSetting may or may n

Strany 44 - • When “Economy” is selected

Coding FieldSelection or rangeFactory setting (bold)Display InfoR2 (4)Configurecleaning function(timer)LC PV = Limitcontactor pH/redox (1)LC °C = Limi

Strany 45 - LC PV R2

1.5 Product structureYou can identify the instrument version by the order code on the nameplate.Sub “codes” are the release codes for Software upgrade

Strany 46 - LC State

Coding FieldSelection or rangeFactory setting (bold)Display InfoR253Enter pre-rinsetime20 s0 ... 999 sWater is used for rinsing.R254Enter cleaningtime

Strany 47

Coding FieldSelection or rangeFactory setting (bold)Display InfoR262Enter setpoint 1(or 2)pH 6.00pH –2.00 ... 16.00Assignment of relay 1 and relay 2fo

Strany 48 - 0.3 R2310

5.7 ServiceCoding FieldSelection or rangeFactory setting (bold)Display InfoSFunction groupSERVICES1Selection oflanguageENG = EnglishGER = DeutschFRA =

Strany 49 - ChemoClean (5)

Coding FieldSelection or rangeFactory setting (bold)Display InfoS9Reset ofinstrument(restore defaultvalues)noSens = sensor dataFacty= factorysettingsF

Strany 50

5.8 E+H Service5.9 InterfacesCoding FieldSelection or rangeFactory setting (bold)Display InfoEFunction groupE+H SERVICEE1Selection ofmoduleContr = con

Strany 51 - Proc.Type

5.10 CalibrationThis function group is used to calibrate thetransmitter. Two types of calibration arepossible:• Calibration by means of measurement in

Strany 52 - 5.7 Service

5.10.1 Wet calibrationFor access to the CALIBRATION menu, please enter Code 22.Coding FieldSelection or rangeFactory setting (bold)Display InfoC(1)Fun

Strany 53 - S.Default

Coding FieldSelection or rangeFactory setting (bold)Display InfoC17Zero point (U_is)is displayedGlass: pH 7.00pH 5.00 ... 9.00Antimony: pH 1.00pH –1.0

Strany 54 - LSG E114

Coding FieldSelection or rangeFactory setting (bold)Display InfoC25Store calibrationresult?yesnonewIf C24 = E xxx, then onlyno or new.If new, return t

Strany 55 - the user is

5.10.2 Numeric calibrationWith the numeric calibration you can enter or correct slope and zero point manually.Coding FieldSelection or rangeFactory se

Strany 56 - ≤ 0.05 and value

2 Safety2.1 Intended applicationThe CPM 223/253 transmitter is a field-testedand reliable transmitter used to determine thepH or redox (ORP) value.The

Strany 57

5.11 OffsetThe settings in the OFFSET function group can be used to calibrate the measurement to a referencemeasurement. This requires a linear shift

Strany 58

6 InterfacesFor instruments equipped with a communication interface, please refer to separate operatinginstructions BA 208C/07/en (HART®) or BA 209C/0

Strany 59 - 59.18 N2

7 Maintenance and troubleshootingMaintenance:• Maintenance means that all measures which will guarantee the safety of operation andreliability of the

Strany 60 - PV Offset

ErrorPossible cause Remedy Equipment need, spare partsFixed, incorrectmeasure value– Sensor not immersed orprotective cap not removedCheck installatio

Strany 61 - 6 Interfaces

ErrorPossible cause Remedy Equipment need, spare partsNo temperature outputsignal– Instrument only has one currentoutputRefer to nameplate for variant

Strany 62

7.2 Troubleshooting using the error messagesDisplay and select error messages by pressing the MINUS key.Errorno.Display MeasuresContakt Error currentA

Strany 63 - Endress+Hauser 61

.Errorno.Display MeasuresContakt Error currentAutomatic startof cleaningFact User Fact User Fact UserE044Stability requirement forcalibration not fulf

Strany 64 - 62 Endress+Hauser

Errorno.Display MeasuresContakt Error currentAutomatic startof cleaningFact User Fact User Fact UserE154Below lower alarmthreshold for periodexceeding

Strany 65 - Endress+Hauser 63

8 Diagnosis and corrective maintenanceDiagnosis• Diagnosis refers to the identification of instrument malfunctions and defects.Corrective maintenance:

Strany 66 - 64 Endress+Hauser

Error Possible cause Tests and / or remedial measures Equipment, spare parts, personnelDisplay showsmeasured values but– value does not changeand /

Strany 67 - Endress+Hauser 65

2.4 Monitoring and safety featuresSafety featuresThe transmitter is protected against externalinfluences and damage by the followingdesign measures:•

Strany 68 - 8.1 Diagnosis

8.2 Corrective maintenance of Liquisys M CPM 2238.2.1 Exploded viewPos. 10Netzteil 100 – 230 VAC230 V100 V115 VSicherungPos. 30 / SteckbrückeNetzspann

Strany 69 - Endress+Hauser 67

8.2.2 Disassembly of CPM 223• Consider potential effects on processwhen removing the instrument from service!• First pull off the terminal block(Pos.

Strany 70 - 223EXPHE.CDR

8.2.4 Spare parts for CPM 223Pos. Designation Name Function Order number10 Power supply unit LSGA 100/115/230 V AC 5150031720 Power supply unit LSGD 2

Strany 71 - 8.2.3 Assembly of CPM 223

8.3 Corrective maintenance Liquisys M CPM 2538.3.1 Exploded view100 VPos. 10Netzteil 100 – 230 V AC230 V115 VSicherungPos. 30 / SteckbrückeNetzspannun

Strany 72 - 8.2.4 Spare parts for CPM 223

8.3.2 Disassembly of CPM 253• Open and remove the connectioncompartment cover (Pos. 420).• Pull out the mains terminal (Pos. 470) tomake the device vo

Strany 73 - 253EXPHE.CDR

8.3.4 Spare parts for CPM 253Pos. Designation Name Function Order number10 Power supply unit LSGA 100/115/230 V AC 5150031720 Power supply unit LSGD 2

Strany 74 - 8.3.3 Assembly of CPM 253

8.3.5 Special case: replacement of central moduleNote:• After controller replacement all theeditable data are reset to thefactory settings.• The seria

Strany 75 - 8.3.4 Spare parts for CPM 253

8.6 Corrective maintenance of measuring system8.6.1 pH/redox transmitterCleaning the front panelTo clean the front panel, only use standardcleaning ag

Strany 76 - 8.4 Spare parts orders

8.6.5 Connecting lines and junction boxesCheck cables and connections for moisture.Moisture reduces the sensor slope. If thedisplay can not be called

Strany 77 - 8.6.4 Liquid KCl supply

9 AccessoriesMounting accessories• Weather protection cover CYY 101For mounting on field housing,for outdoor installation.Dimensions (H × W × D):320 ×

Strany 78

3 InstallationThe following procedure should be followed for a complete measuring system installation:• Installation or attachment of transmitter(see

Strany 79 - 9 Accessories

Assemblies• DipFit W CPA 111Immersion assembly for pH andredox electrodes with bayonet system,applicable for up to 4 bar.Technical Information TI 112C

Strany 80

10 Technical dataGeneral specificationsManufacturer Endress+HauserProduct designation Liquisys M CPM 223Liquisys M CPM 253InputMeasured variables pH,

Strany 81 - 10 Technical data

Output pH signal outputCurrent range 0 / 4 ... 20 mA, galvanically separated; error current 2.4 / 22 mAMax. resolution 700 digits / mALoad max. 500 ΩO

Strany 82 - 80 Endress+Hauser

Accuracy pH measurementReference temperature +25 °CMeasured value resolution 0.01 pHDisplay deviation1max. 0.5 % of measuring rangeRepeatability1max.

Strany 83 - Endress+Hauser 81

11 AppendixVZNV2N2V3N3V4N4 N4V1N1CC13 C15C11 C12C14C16C25C23C24C22C32C34C36C35C31C33C21AA1A2A3A4A5 A6BB1B2B3B4B5FF1F2F3F4F5F6OO1O2O3 (3)O3 (2)O3 (1)O3

Strany 84 - 11 Appendix

C19C17C18F8F7F9O335O336O334Z7Feedforward control =1at0 … 100%50%Store calibrationresults; no; newyesZero point is displayedGlass5.00...9.00 pHAntimon-

Strany 85 - Endress+Hauser 83

RFunction groupRELAYSelect contact to beconfigurdRel2; Rel3; Rel4;Rel1;R1R2 (6)R2 (5)R2 (4)R2 (3)R2 (2)R2 (1)FunctionSwitch off or on;onR2 (2)offR221E

Strany 86 - 84 Endress+Hauser

Dropout delayentry0...2000 s0sR225Entry of alarm threshold(as an absolute value)150.0 °C-20.0...+150.0 °CR226Entry of derivative actiontime Tv(0.0 = n

Strany 87 - Endress+Hauser 85

12 Index!0 ...18,2122 ...219999 ...18,21AA-Function group . . . . .

Strany 88 - 12 Index

LLED indicator . . . . . . . . . . . . . . . . . . . . . . . . 17Limit contactor . . . . . . . . . . . . . . . . . . . . 38, 43Limit contactor for con

Strany 89 - PM253E12.CHP

3.2 DimensionsNote:There is a hole in the punching forPg 16 cable entry. It serves as apressure balance during air freightdispatching. Make sure that

Strany 90

SS-Function group . . . . . . . . . . . . . . . . . . . . . 50Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 5Safety features . .

Strany 91 - Declaration of contamination

More information about services and repairs:www.services.endress.comDeclaration of contaminationDear customer,Because of legal determinations and for

Strany 92 - Printed in Germany / CV5 / DT