8.3.5 Special case: replacement of central module
Note:
• After controller replacement all the
editable data are reset to the
factory settings.
• The serial number can only be
entered – and only once – in the
case of a new module from the
factory with serial number 0000!
Make sure that your entry is
correct before confirming with
ENTER!
Entry of an incorrect code will
prevent the enhanced functions
from being enabled. An incorrect
serial number can only be
corrected at the factory.
Proceed as described below after central
module replacement:
• If possible, record the user settings of the
instrument, e.g.:
– Calibration data
– pH / mV and temperature current
assignments
– Relay function selections
– Limit / controller settings
– Cleaning settings
– Monitoring functions
– Interface parameters
• Dismantle the device as described in
Chap. 8.2.2 or 8.3.2 .
• Refer to the part no. of the central module
to determine whether the new module has
the same part no. as the old one.
• Reassemble the device with the new
module, as described in chap. 8.2.3 or
8.3.3.
• Put the device into operation again and
check the fundamental function (e.g.
measured value and temperature display,
operability via keyboard).
• Enter the instrument serial number:
– Read the serial number (“user-no.”)ofthe
device from the nameplate.
– Enter this number in the fields E115 (year,
one-digit), E116 (month, one-digit),
E117 (sequence number, four-digit).
– Field E118 displays the complete number
for verification; acknowledge with ENTER
or abort and re-enter.
• Check if the S packet has been activated
(e. g. by calling up the function group
CHECK / P code) or the Chemoclean
function.
• Restore the user settings of the instrument.
8.4 Spare parts orders
Spare parts are to be ordered from your local
E+H Sales Agency. The address is on the
back cover of these operating instructions.
Use the order numbers listed in Chapters
8.2.4 and 8.3.4
To be on the safe side, you should always
specify the following data with spare parts
orders:
– Instrument order code (order code)
– Serial number (user-no.)
– Software version where available
Refer to the nameplate for the order code and
serial number.
The software version is displayed in field
E111 when the instrument processor system
is functional.
8.5 Service equipment “Optoscope” with “Scopeware”
The Optoscope together with the
“Scopeware” software offers the following
possibilities, without having to remove or
open the Liquisys M and without galvanic
connection to the device:
• Documentation of the instrument settings in
conjunction with Commuwin II
• Software update by the service technician
• Upload/download a hex dump to duplicate
configurations.
The optoscope serves as an interface
between the Liquisys M and PC / laptop. The
information exchange takes place via the
optical interface on the Liquisys M and via an
RS 232 interface on the PC / laptop.
Handling and operation are described in the
optoscope operating instructions. The
user-friendly Windows software required for
the PC or laptop is supplied with the
optoscope.
The optoscope is supplied in a sturdy case
with all the accessories required.
Order number of optoscope: 51500650
Diagnosis and corrective maintenance Liquisys M CPM 223 / 253
74 Endress+Hauser